BEIJING HOPELASER TECHNOLOGY CO LTD

Best quality ,best price, best service

WHATSAPP/ VIBER/WECHAT :+8613811977881

Skype :BEIJINGMEDICALBEAUTY

 

Home
Products
About Us
Factory Tour
Quality Control
Contact Us
Request A Quote
Home

BEIJING HOPELASER TECHNOLOGY CO LTD Quality Control

Certification
Good quality HIFU Machine for sales
Good quality HIFU Machine for sales
Customer Reviews
whatsapp: 008613121716589

—— whatsapp

I'm Online Chat Now
QC Profile


E light equipment maintenance guide series


No. fault failure analysis or solutions
1, No power or bad energy E to E; E to the output line or bad, or loose Leakage of E
2, to the power transistor is not installed on the good 2233, 2611
3, instruments showed good head not in contact with the skin contact surface and the skin not exposed to good
4, LCD display, go to "pre-ignition state" LCD signal short-circuit, the treatment switch bad.
5, liquid crystal display "Please release the handle switch" LCD signal short-circuit or a bad switch handle
6, LCD display apparatus without water or power cable 15V bad bad
7, temperature sensor bad 18B20; such as re-boot properly, then the two-way absorption of bad management
LCD 

8,count no light or power in the integrated circuit 358, 3525 or 4094 / FET FR460, diode 10150 or motherboard bad control board 4-400B

9, no lighting equipment bald simmer plate / diode BY359 / high SCR / JW100K resistance / bad IC 3525
10, jump relay contactor auxiliary contact bad; bad simmer plate; bad LCD panel

Exchanges and contacts the lamp 11 does not pull bad simmer plate
12, LCD blue screen bad, crystal
13, LCD blue screen, buzzer alarm LCD main chip bad


The Common problem (faults) and solution with intense pulsed light(IPL) treatment


Intense pulsed light therapy machine (using the 800W and 2000W power supply)
1, cant light the lamp of ipl handle;
Failure analysis:
(1)ipl handpiece(ipl lamp) bad
(2) bad power supply
(3)simmer board failure
Solution: Try to replace a new ipl handle (IPL lamp), in case the problem
is still same , replace the ipl power


2, no effect after treatment, the treatment energy is too low ;
Failure analysis:

(1) ipl handle used for a long time, energy consumption of ipl xenon lamp.
(2) power supply or LCD is not working properly, the voltage of the capacitor is low.
Solution:

(1) explain the situation to the customer, make the appropriate adjustments to the operating parameters, serious energy consumption of IPL xenon lamp, the longevity of ipl lamp is use up .
Need to replace a new ipl lamp or ipl handle
(2) replace the power supply or LCD.


3, unstable energy or energy become weak gradually , and finally no light shot;
Failure analysis:

(1) .Each time a larger difference between the discharge interval, the pulse interval is shorter.
(2) Power , LCD control panel not work
Solution:

(1) to keep each discharge interval of 3 seconds.
(2) replace the power supply or LCD.


4, cant turn ipl machine on ;
Failure analysis:

(1) bad power cord
(2) Emergency switch does not bounce
(3) key switch bad
(4) Bad fuse (burn up fuse)
(5) the air switch is disconnected
Solution:

(1) the replacement a new power cord
(2) Emergency switch bounce
(3) Replace key switch
(4) Replace the fuse
(5) the air switch to closed state


5, ipl shot automatically after simmer ;
Failure Analysis: IGBT bad
Solution: Replace the broken IGBT or replace the power , need professional maintenance(Beijing ).


6, counting , but no light (no ipl shot );
Failure Analysis: LCD panel or power board bad
Solution: replace the power supply or LCD panel


7, ipl handle heat
Failure analysis:

(1) there is no water flow , or no water added in the water tank
(2) long working hours


Solution:

(1) adding water to the overflow hole, place the ipl handle on the ground, so as to make sure that the water circulation is good enough
(2) Do not work too long. For a lot of clients , pay attention to changing the water, first water tank release 2 / 3, keep 1 / 3, re-add water to overflow hole, rest 30-40 minutes before work.

 

8, touch screen displays only half or blue screen;
Failure Analysis: Bad LCD
Solution: Replace LCD

 

9, LCD screen blank, pump and fan working properly;
Failure analysis:

(1) LCD bad
(2) no power supply for LCD screen
Solution:

(1) the replacement of the LCD
(2) check the cable, to ensure that the LCD's main power supply.

 

10, automatic lighting(ipl lamp light automatically );
Failure analysis: (1) LCD bad
Solution: (2) the replacement of the LCD

 

11, showing no water circulation or no water;
Failure analysis:

(1) bad water pump
(2) the short circuit for water probe
(3) LCD panel bad or power supply bad
Solution:

(1) take out the ipl handle, if there is flowing out, indicating that the pump normally, if there is no water in the ipl handle . water pump bad,
You can also hear the sound of water flow on the ipl handle when you turn on machine
(2) connect water probe well.
(3) the replacement of the LCD panel or the power board

 

12, ipl handle is not cooling, not shot, room temperature;
Failure analysis:

(1) switching power supply bad (12V DV switch power)
(2) Refrigeration signal cable is not plugged in
(3) cooling chip bad
Solution:

(1) the replacement of switching power supply
(2) connect the refrigeration signal cable well.


Common problem (faults)and solutions for laser treatment


1, laser lamp simmer ok , can shot laser, but without focus.
Reasons:

(1) Laser rod leakage
(2) front mirror was breakdown. The Front and back tips of the focusing mirror was opposite .
(3) laser handle burn up,
(4) un-focus
Solution:

(1) replace laser rods
(2) Replace the front mirror, adjust the the positive and negative of focusing mirror
(3) replace the gun
(4) dimming

2, laser lamp light, but no shot
Reasons:

(1) does not count, no shot
(2) counting , not shot
Solution:

(1) the replacement of foot switch
(2) replacement power supply or LCD

3, ipl lamp cant light
Reasons:

(1) The lamp is broken or the power supply is broken
(2) bad LCD panel
Solution:

(1) when you simmer, if hearing the sound of pa pa pa, the lamp is broken, replace the lamp; if not , it is a problem with power supply
(2) the replacement of the LCD panel

4, cant turn on laser machine
Reasons:

(1) broken power cord
(2) The emergency switch does not bounce
(3) emergency switch or key switch broken
(4) broken laser power supply
(5) burn up fuse
Solution:

(1) the replacement power cord
(2) the emergency switch bounce
(3) replacement of the emergency or key switch
(4) replace the power
(5) Replace the fuse

5, no display after turn on laser machine, but the water pump work
Reasons:

(1) broken fuse for laser power
(2) LCD bad
(3) cable disconnected
(4) broken power supply
Solution:

(1) Replacement Power fuse
(2) the replacement of the LCD panel
(3) Check the cable connecting well or not
(4) maintenance for power

6, turn on laser machine properly, the water does not flow
Reasons:

(1) there are air in the water pipes(water tube)
(2) bad water pump
Solution:

(1) hold up the laser handle (more than machine ), and then lower (less than machine); repeatedly turn on and off machine ;
(2) replace the water pump

7, laser handle leakage
Reasons:

(1) long time working,
(2) the laser lamp and laser rod broken
 
Solution:

(1) shorter working hours
(2) the replacement of laser lamps or the laser rod
  

8, shot automatically after the laser lamp simmer successfully
Reasons:

(1) the footswitch broken or short circuit )
(2) bad LCD panel
(3) power bad
Solution:

(1) the replacement of foot switch
(2) the replacement of the LCD panel
(3) the replacement of power

9, the replay jump when simmer or working .
Reasons:

(1) laser lamp the wrong connect with Positive and anode,
(2) lamp cable disconnected well
(3) The lamp is broken
(4) Bad power 
Solution:

(1) exchange positive and anode position ,
(2) Check laser lamp cable if it is connected well or not
(3) for light
(4) maintenance of power

 

Certification
  • BEIJING HOPELASER TECHNOLOGY CO LTD

    Standard:EC

    Number:AGC1085099GZ80E

    Issue Date:2009-04-17

  • BEIJING HOPELASER TECHNOLOGY CO LTD

    Standard:Medical CE(medical beauty)

    Number:HD600251480001

    Issue Date:2009-06-24

  • BEIJING HOPELASER TECHNOLOGY CO LTD

    Standard:CE1023

    Number:080736QS/NB

    Issue Date:2009-07-08

    Scope/Range:RF beauty equipment

  • BEIJING HOPELASER TECHNOLOGY CO LTD

    Standard:CE0928

    Issue Date:2009-12-08

    Scope/Range:cavitatation rf machine

Contact Details
BEIJING HOPELASER TECHNOLOGY CO LTD

Contact Person: Mr. Tim

Tel: 86-10-89249884

Fax: 86-10-89249884-888

Send your inquiry directly to us (0 / 3000)